Precipitation
Transmitter
The instrument is designed to measure the height, quantity and the intensity of the precipitation striking the surface of the earth. The measuring principle, tipping bucket, is basing on the description "Guide to Meteorological Instruments No 8" of the WMO (World Meteorological Organization). The precipitation, collected by the collecting surface and the collecting funnel, is conducted into a tipping-bucket. The tipping bucket consists of two bucket-compartments. Is one of these compartments filled with water it tips over, and the water drains off. Meanwhile subsequent rain falls into the newly positioned upper compartment. The tipping movement is detected by a Reed-contactor, and a connected electronics, and produces a respective output signal.
There are two outputs available:
1. Analogue output for the output of the precipitation sum as voltage- or current value.
2. Pulse output for the output of single precipitation meter pulse.
The electronics of the precipitation
transmitter is equipped with a linearising system. The linearising procedure
is basing on a precipitation-/intensity-dependent pulse number correction for
the range from approx. 0,5...7 mm/min. In our laboratory each instrument
is calibrated within the intensity range of 0...7 mm/min with a water quantity
of 200cm³ (=10 mm precipitation height).
Precipitation transmitters with built-in heating (5.4033.35.xxx) liquefy snow
or hail and consequently, are suitable for winter use. All parts are corrosion-resistant.
The casing consists of stainless steel (V2A).
|
|
|
|||
|---|---|---|---|---|
|
5.4033.35.xxx
|
Heating | 70 W; 24 V AC/DC | ||
|
36.xxx
|
no | |||
|
.040
|
Output/Analogue | 0 ... 20 mA (<500 ohm) | ||
|
.041
|
4 ... 20 mA (<500 ohm) | |||
|
.073
|
0 ... 5 V | |||
|
.061
|
0 ... 10 V | |||
| Collecting surface | 200 cm² | |||
| Resolution | 0,1 mm | |||
| Measuring range | max. 7 mm/min. | |||
| Measuring principle | tipping bucket | |||
| Supply voltage | 24 V AC/DC or | |||
| (w/o
heating) 10 ... 28 V DC 14 ... 28 V DC (10V-output) |
||||
| Ambient temperature | -25... +60°C with heating | |||
| 0... +60°C w/o heating | ||||
| Mounting | on stand pipe (1 ½") | |||
| Weight | 3,3 kg | |||