Precipitation
TransmitterGeneral Information
This instrument is designed to measure precipitation striking the surface of the earth. The precipitation strikes a collector with a surface area of 200 cm² and then travels via a run-in sieve into a tipping scale. When a certain amount of water has gathered, the scale tips and empties, while the other half of the scale is ready to receive water again. This tipping procedure is detected by a reed switch and is made available as a pulse for an electrical Signal.
The signal is prepared and emitted on the digital output (optocoupler) as a pulse. Simultaneously the input signal is converted in an analog value change and made available on the analog output as a current or voltage value. (See Models Available) The amount of the analog value change depends on the measuring range selected. There are different measuring ranges available (see Selection of Measuring Range). Selection is made by means of a DIP switch and thus can be adjusted by the user to the local conditions. The analog output is set automatically to the initial value (for example 0V..., 4mA..., 0mA...) when the measuring range transmission value has been reached.
Resetting the analog output to the initial value can also be done with a RESET-impulse on the reset-input (at any time). The Precipitation Transmitter is equipped with an electronically regulated heater for winter operation.
Models Available
|
|
|
|
|
| 5.4033.30.040 | 0 ... 20 mA (max. 500 O ) |
|
|
| 5.4033.30.041 |
|
4 ... 20 mA (max. 500 O ) |
|
| 5.4033.30.061 |
|
0 ... 10V |
|
| 5.4033.31.040 |
|
0 ... 20 mA (max. 500/300 W) |
|
| 5.4033.31.041 | 4 ... 20 mA (max. 500/300 W) |
|
|
| 5.4033.31.061 |
|
0 ... 10 V |
|
The factory setting for the measuring
range is 0 ... 10 mm.
Technical Data
|
|
|
|
| Heating |
|
|
| Switch-on temperature |
|
|
| Hysterese |
|
|
| Heating power |
|
|
| Ambient temperature |
|
|
| Measuring system | Tipping bucket | |
| Resolution | 0.1 mm | |
| max. Intensity | 7 mm/Minutes | |
| analog Measuring range | adjustable ( DIP Switch) | |
| Measuring range 1 | 0 ... 10 mm Precipitation | |
| Measuring range 2 | 0 ... 25 mm Precipitation | |
| Resolution | 0.1 mm | |
| Measuring range 3 | 0 ... 20 mm Precipitation | |
| Measuring range 4 | 0 ... 50 mm Precipitation | |
| Resolution | 0.2 mm | |
| analog registration | with automatic overflow | |
| with external reset | ||
| Reset input | Optocoupler | |
| Input voltage | 5 ... 30 V | |
| Pulse | ³ 20 msec | |
| Measured value output | ||
| 1. Digital | ||
| Optocoupler | 1 Pulse/Tipping bucket | |
| Pulse length | >= 100 msec | |
| U (max.) | 30 V | |
| 2. Analog | ||
| Voltage output | 0 ... 10 V (corresponds to...s. Measuring range) | |
| Current output | 0 ... 20 mA (corresponds to s. Measuring range) | |
| 4 ... 20 mA | ||
| Non linearity | ± 0,2 % (-30°C ... 60°C) | |
| Offset error | < 0,3 % (-30°C ... 60°C) | |
| load | 500 W Power supply 24 V AC/DC | |
| 300 W Power supply 10 ... 14 V DC | ||
| Power supply | 1. 24 V AC/DC (±15%) | |
| 2. 10 ... 18 V DC | ||
| Power consumption | approx. 20 mA (without heating) | |