Heat Flux Plate HFP01
Part number: 7.1417.10.000
-
Description
HFP01 measures heat flux through the object in which it is incorporated or on which it is
mounted, in W/m2
The sensor in HFP01 is a thermopile. This thermopile measures the temperature difference across the ceramicsplastic composite body of HFP01. A thermopile is a passive sensor; it does not require power. Using HFP01 is easy. It can be connected directly to commonly used data logging systems. The heat flux in W/m² is calculated by dividing the HFP01 output, a small voltage, by the sensitivity. The sensitivity is provided with HFP01 on its calibration certificate.
A typical measurement location is equipped with 2 or more sensors. HFP01 is the world’s most popular sensor for heat flux measurement in the soil as well as through walls and building envelopes.
Unique features and benefits
• low thermal resistance (essential for use on walls and windows)
• large guard area (required by the ISO 9869 standard)
• low electrical resistance (low pickup of electrical noise)
• high sensitivity (good signal to noise ratio in low-flux environments such as buildings)
• robustness, including a strong cable
• IP protection class: IP67 (essential for outdoor application)Specification
Heat fluxMeasuring range-2000 ... 2000 W/m²Sensitivity60 µV/W/m²Typ. signal output-120 ... +120 mVSensor typeThermocouplesImpedance1 ... 4 êAccuracy±3 %Time response4 minTemperature coefficient0.1 %/KGeneralAmbient temp.-30 ... +70 °CElectr. connection5 m cableProtectionIP 67DimensionØ 80 x 5.4 mmWeight0.2 kgVersions
No other versions of this product are available.Accessories
No accessories are available for this product.Further information